PT.Mora Anugerah Berkat adalah perusahaan penyedia alat – alat laboratorium, alat laboratorium yang kami sediakan meliputi : general laboratorium equipment,medical equipment (alat kesehatan), environmental (alat uji lingkungan), chemical/reagent consumable,mineral/civil material testing dan kalibrasi.

How to Use Zahn Cup Test

 




How to Use Zahn Cup Test

Description 

BGD 126 Zahn Cups can be used to quickly measure the viscosity of liquids such as Paint, Ink, Varnishes, Syrup and Oil. They are produced according with ASTM D 816, ASTM D 1084,and ASM D 4212. The Stainless Steel Cups are molded precisely and the orifices are precision drilled using high-speed CNC drilling.


Each cup has a 12-inch loop handle to allow the cup to be dipped by hand into a liquid container.  At the center of this handle is a finger-ring for holding the cup in a vertical position during use. Their orifice diameters are set at the factory for appropriate results with applicable NIST traceable Newtonian oils.


It is done by filling up the volume of Zahn Cup with the required liquid and then placing an index finger over the orifice to stop any flow out of the liquid. A ring at the Zahn Cup handle allows the cup to be hanged perpendicularly(still with the finger stopping any liquid flow out). Prepare a Digital stopwatch and set zero. Immediately start the stopwatch when the finger is released. As soon as there is a first “break” of the flow, stop the stopwatch. The second shown is the time of the liquid flow, which is also the flow viscosity of that liquid.


Every Biuged’s Zahn cup has its own coefficient “K”, which is marked on the cup body. 

Fact Viscosity = Testing Viscosity × K


Main Technical Parameters

  • Volume of Cup :43ml - 49ml

  • Length of Handle:40mm ± 0.1mm

  • Height of Cup:58mm ± 0.1mm

  • Measurement Temperature:25℃±1℃

  • Come with calibration certificate 

  • Details of each Zahn Cups

  • Click Here : Zahn Cup / Zahn Viscosity Cup Brochure


How to Use Zahn Cup Test


Details /
Zahn Cup
Zahn Cup No.1Zahn Cup No.2Zahn Cup No.3
Orifice
(mm/inches)
1.98 / 0.082.74 / 0.113.76 / 0.15
Flow Time
Range (sec)
33.5 - 8020 - 8020 - 75
Measurement
Range (CST)
5 - 5621 - 231146 - 848
ApplicationsThin-bodied materialsClears, lacquers, enamels,
press-side adjustment
of flexographic inks
Manufacturing of
flexographic inks
Ordering
Information
BGD 126/1BGD 126/2BGD 126/3
Zahn Cup / Zahn Viscosity Cup
Details /
Zahn Cup
Zahn Cup No.4Zahn Cup No.5
Orifice
(mm/inches)
4.27 / 0.175.28 / 0.21
Flow Time
Range (sec)
20 - 8020 - 80
Measurement
Range (CST)
222 - 1110460 - 1840
ApplicationsMore viscous
paints and inks
Silk screen inks
Ordering
Information
BGD 126/4BGD 126/5


=========================================================
How to Use Zahn Cup Test
=========================================================

Zahn Viscosity Cup​

The Zahn Viscosity Cup is used to measure and adjust the viscosity of ink. The viscosity will be measured by the time the ink took to flow down from the orifice.


Cup NoViscosity※Time
120 to 85 cP40 to 85 s
230 to 170 cP20 to 70 s
3170 to 550 cP15 to 60 s
4200 to 900 cP20 to 65 s
5250 to 1,200 cP15 to 60 s

How to Use Zahn Cup Test

Untuk informasi,pertanyaan dan pemesanan silahkan hubungi kami di : 
  • PT. Mora Anugerah Berkat [M.A.B]
  • Click Here : Zahn Cup / Zahn Viscosity Cup Brochure

  • Alamat : Jl.Taman Asri Blok M No.36 Kel.Cipadu Kec.Larangan, Kota Tangerang - Provinsi Banten 
  • Phone : 021 7345.7064 
  • Mobile : 0812.8140.0749 / 0812.8770.5470
  • Chat : Whats App, Line, WeChat, Telegram 
  • Website www.moraanugerah.id -
  •                           www.moraanugerah.com
  • Email : info@moraanugerah.com -
  •                   marketing@moraanugerah.com -  
  •                   marketing_div@moraanugerah.com       


0 Komentar: