- Dapatkan link
- X
- Aplikasi Lainnya
- Dapatkan link
- X
- Aplikasi Lainnya
Gas Cooler Series RC 1.1
In the chemical industry, petrochemistry or biochemistry, reliable process control relies on prompt and exact determination of the operating parameters. Here, gas analysis is the key for safe and efficient control of process flows, environmental protection and quality assurance. This benefits controlling flue gas emission in power stations or exhaust gas analysis in automotive engineering, as well as the efficient control of air separators or sterile production and packaging in the food industry. Many of the analysis processes used in these fields require extracting the sample gas. This inevitably also extracts process-related contamination such as particles or moisture. These in turn can impact the measurement results or damage the measuring cells. The sample gas must therefore be conditioned before entering the analyser. The RC 1.1 is a compressor sample gas cooler for up to 2 gas paths and is an essential component for sophisticated analysis systems
GAS COOLER SERIES RC 1.1
- Compact design: Pre-installed and ready to connect
- Choose from one or two gas paths
- Heat exchanger made from stainless steel, Duran glass and PVDF
- Adjustable outlet dew point and alarm thresholds
- Cooling block temperature display
- Status display and output
- Rated cooling power 360 kJ/h
- Constant dew point stability ± 0.1 °C
- Optional 4 - 20 mA or Modbus RTU signal output
- Moisture detector, filter and condensate pump optional
- Successor to the EGK 1/2 and EGK 1SD
- Click Here : Sample Gas Cooler RC 1.1 Brochure
- Click Here : Gas Cooler Series TC-MIDI Brochure
Gas Cooler Series RC 1.1
| Gas Cooler Technical Data | GAS COOLER SERIES RC 1.1 |
| Ready for Operation | after max. 15 minutes |
| Rated Cooling Capacity (at 25°C) | 360 kJ/h |
| Ambient Temperature | 5 °C to 50 °C |
| Gas Outlet Dew Temperature,Preset | approx. 5 °C |
| Dew Point Fluctuations Static: | ± 0.1 K |
| In the Entire Specification Range: | ± 1.5 K |
| IP rating | IP 20 |
| Housing | Stainless steel |
| Packaging Dimensions | Approx. 420 x 440 x 350 mm |
| Weight Incl. Heat Exchanger | approx. 15 kg |
| Electric Supply | 115 V, 60 Hz or 230 V, 50/60 Hz ± 5% Plug per DIN EN 175301-803 |
| Electrical Data | 230 V 115 V |
| Typical Power Input : | 296 VA 302 VA |
| Max. Operating Current : | 2 A 3.5 A |
| Alarm Output Switching Connection | max. 250 V, 2 A, 50 VA Connector per DIN EN 175301-803 |
| Installation | stand-alone or wall-mounted |
| Technical Data Peristaltic Pumps CPsingle / CPdouble | |
| Flow Rate | 0.3 L/h (50 Hz) / 0.36 L/h (60 Hz) with standard hose |
| Vacuum Inlet | max. 0.8 bar |
| Pressure Inlet | max. 1 bar |
| Outlet Pressure | 1 bar |
| Hose | 4 x 1.6 mm |
| Condensate Outlet | Hose nipple Ø6 mm Screw connection 4/6 (metric), 1/6"-1/4" (US) |
| Protection Class : | IP 40 |
| Materials Hose : | Norprene (Standard), Marprene,Fluran |
| Connections: | PVDF |
| Analogue Output Cooler Temperature Technical Data | |
| Signal | 4-20 mA or 2-10 Vcorresponds to -20 °C to +60 °C cooler temperature |
| Connection | M12x1 Plug, DIN EN 61076-2-101 |
| Technical Data Filter AGF-PV-30-F2-L | |
| Ambient Temperature | 3 °C to 100 °C |
| Max. Operating Pressure with Filter | 2 bar |
| Filter surface | 125 cm2 |
| Filter mesh | 2 μm |
| Dead volume | 108 ml |
| Materials | |
| Filter: | PVDF, Duran Glass (parts in contact with media) |
| Seal | Viton |
| Filter element | sintered PTFE |
| Technical Data FF-3-N Moisture Detector | |
| Ambient Temperature | 3 °C to 50 °C |
| Max. Operating Pressure with FF-3N | 2 bar |
| Material | PVDF, PTFE, epoxy resin, stainless steel 1.4571, 1.4576 |
Many gas analysis processes require extracting sample gas from the process. This also extracts process-related contamination such as particles or moisture. These can impact the measurement results or damage the measuring cells. The sample gas must therefore be conditioned before entering the analyser. The sample gas cooler reduces the gas temperature to below the dew point for this purpose, causing moisture to drop out, which is then discharged as condensate.
In addition to the status output to monitor the sample gas cooler function, we offer an optional 4 – 20 mA analog output or digital interface. The process control can access the process and diagnostic data via the Modbus RTU interface as well as configure the device settings.
The RC 1.1 is a compressor sample gas cooler designed for high cooling output and high ambient temperatures. Depending on the application, it can be configured with one or two gas paths. This makes it a vital cooling component in sophisticated analysis systems.
Gas Cooler Series RC 1.1
| Heat Exchanger | TS TS-I2) | TG TG | TV-SS TVSS-I 2) |
| Version / Material | Stainless Steel | Glass | PVDF |
| Flow Rate vmax 1) | 530 L/h | 280 L/h | 155 L/h |
| Inlet Dew Point τe,max1) | 80 °C | 80 °C | 65 °C |
| Gas Inlet Temperature ϑG,max1) | 180 °C | 140 °C | 140 °C |
| Max. Cooling Capacity Qmax | 450 kJ/h | 230 kJ/h | 120 kJ/h |
| Gas Pressure pmax | 160 bar | 3 bar | 3 bar |
| Pressure drop Δp (v=150 L/h) | 8 mbar | 8 mbar | 8 mbar |
| Dead Volume Vtot | 69 ml | 48 ml | 129 ml |
| Gas Connections (metric) | G1/4 | GL14 (6mm)4) | DN 4/6 |
| Gas Connections (US) | NPT 1/4" | GL14(1/4")4) | 1/4"-1/6" |
| Condensate Out Connections(metric) | G3/8 | GL25(12mm)4) | G3/8 |
| Condensate Out Connections (US) | NPT 3/8" | GL25(1/2")4) | NPT 3/8" |
| Heat Exchanger | DTS(DTS-63)) DTSI(DTS6-I3))2) | DTG DTG | DTV3)DTV-I2)3) |
| Version / Material | Stainless steel | Glass | PVDF |
| Flow Rate vmax 1) | 2 x 250 L/h | 2 x 140 L/h | 2 x 115 L/h |
| Inlet Dew Point τe,max1) | 80 °C | 65 °C | 65 °C |
| Gas Inlet Temperature ϑG,max1) | 180 °C | 140 °C | 140 °C |
| Max. Cooling Capacity Qmax | 450 kJ/h | 230 kJ/h | 185 kJ/h |
| Gas Pressure pmax | 25 bar | 3 bar | 2 bar |
| Pressure drop Δp (v=150 L/h) | 5 mbar each | 5 mbar each | 15 mbar each |
| Dead Volume Vtot | 28 / 25 ml | 28 / 25 ml | 21 / 21 ml |
| Gas Connections (metric) | 6 mm tube | GL14(6mm) 4) | DN 4/6 |
| Gas Connections (US) | 1/4" tube | GL14(1/4")4) | 1/4"-1/6" |
| Condensate Out Connections(metric) | Tube 10 mm (6 mm) | GL18(10mm)4) | DN 5/8 |
| Condensate Out Connections (US) | Tube 3/8" (1/4") | GL18(3/8")4) | 3/16"-5/16" |
Gas Cooler Series RC 1.1
- PT. Mora Anugerah Berkat (M.A.B)
- Click Here : Sample Gas Cooler RC 1.1 Brochure
- Click Here : Gas Cooler Series TC-MIDI Brochure
- Alamat : Jl.Taman Asri Blok M No.36 Kel.Cipadu Kec.Larangan,Kota Tangerang - Provinsi Banten
- Phone : 021 7345.7064
- Mobile/HP : 0812.8140.0749 / 0812.8770.5470
- Chat : Whats App, Line, WeChat, Telegram
- Website : www.moraanugerah.id - www.moraanugerah.com
- Email : info@moraanugerah.com - marketing@moraanugerah.com - marketing_div@moraanugerah.com
alat gas cooler
gas cooler
gas cooler rc 1.1 buhler
harga gas cooler rc 1.1 buhler
harga jual gas cooler buhler
jual gas cooler rc 1.1 buhler
spesifikasi gas cooler rc 1.1 buhler
supplier gas cooler
- Dapatkan link
- X
- Aplikasi Lainnya


Komentar
Posting Komentar