IKA Bomb Calorimeter C 1
The C 1 is a tiny giant. It takes up hardly any space in the laboratory, yet offers a high degree of automation. Where there is no need for compliance with standards, however calorific values still need to be determined or processes tested with a high degree of accuracy, this calorimeter makes life so much easier.It works according to the isoperibolic principle and with a static jacket. This makes the C 1 a strong partner for industrial and testing laboratories.
It is especially popular in: the waste management industry, the foodstuffs and animal feed industry as well as in the construction materials sector.
C1 Functions
The light, easy-to-attach combustion chamber replaces the decomposition vessel with a screw cap.
The unit functions fully automatically, which makes it particularly user-friendly
> Automatic oxygen filling, venting and flushing
> Automatic ignition with a fixed ignition wire and ignition energy specification
> Automatic water filling and drainage
> Works with a chiller (RC 2 basic) or with an existing cold water connection and a heater (C 1.20, accessory)
> Interfaces for PC (USB-B), printer (serial interface), scales (serial interface)
Your Laboratory, our C 1 Package
What kind of laboratory do you have? With us, you will certainly find the right C 1 package. For example, if you only have a cold-water connection and your samples have a high halogen content, we recommend the C 1 package 3/12. The therein contained water heater guarantees the temperature required
for its operation and ensures reproducible results. The C 1.12 decomposition vessel, included in the delivery, provides the necessary corrosion protection for chlorinated samples.
What do you need? Just get in contact with us!
Packages including combustion chamber C 1.10
C 1 PACKAGE 1/10 | Ident. No. 0010002412
C 1 Calorimeter including RC 2 basic Circulating chiller
C 1 PACKAGE 2/10 | Ident. No. 0010002413
C 1 Calorimeter
C 1 PACKAGE 3/10 | Ident. No. 0010002414
C 1 Calorimeter including C 1.20 Heater
Packages including combustion chamber C 1.12
C 1 PACKAGE 1/12 | Ident. No. 0010002415
C 1 Calorimeter including RC 2 basic Circulating chiller,halogen resistant
C 1 PACKAGE 2/12 | Ident. No. 0010002416
C 1 Calorimeter, halogen resistant
C 1 PACKAGE 3/12 | Ident. No. 0010002417
C 1 Calorimeter including C 1.20 Heater,halogen resistant
Technical Data C 1 Calorimeter | |
---|---|
TECHNICAL DATA | |
Measuring range max | 40.000 J / 9,560 cal |
Temperature measurement resolution | 0.0001 K |
Oxygen operating pressure max. | 30 bar |
Display | TFT |
Multifunctional knob (turn / push) | Yes |
Reproducibility static jacket (1g benzoic acid NBS39i) | 0.15 % RSD |
Measurements per hour | 4 |
Jacket control | Static, dry |
Start temperature setting | 2 possibilities: 22 °C or 30 °C |
Number of possible decomposition vessels per device | Up to 2 |
INTERFACES | |
Interface PC | USB-B |
Interface Printer / Balance | 9 pin (M) RS 232 serial |
AUTOMATIC FUNCTIONS | |
Automatic oxygen filling, degasing and flushing | Yes |
Automatic water filling and draining | Yes |
"Automatic combustion with ignition wire and combustion | |
energy determination with every sample run" | Yes |
COOLING WITH RC 2 BASIC CHILLER | |
Cooling medium temperature min. | 18 °C |
Cooling medium temperature max | 29 °C |
Cooling medium permissible operating pressure | 1.5 bar |
Cooling medium | Tap water |
Type of cooling | Flow |
Flow rate min | 50 l/h |
Flow rate max | 60 l/h |
Recommended flow rate at 18 °C | 55 l/h |
GENERAL DATA | |
Languages | DE, EN, FR, ES, CN, RU, PO, IT |
Weight | 15 kg |
Dimensions open (W × H × D) | 290 × 400 × 300 mm |
Dimensions closed (W × H × D) | 290 × 280 × 300 mm |
Permissible ambient temperature | 20 - 25 °C |
Permissible relative humidity | 80% |
Voltage | 100 - 240 V |
Frequency | 50 / 60 Hz |
Power input | 150 W |
Untuk mengetahui model/type dan spesifikasi Bomb Calorimeter Click link tautan yang ada di bawah ini :